Instrument description:
Gas chromatograph Shimadzu GP2010, carrier gas: He. Autosampler AOC 6000 (PAL 3) with the capability for liquid sample analysis and HeadSpace analysis.

Available columns:

Supplier Type l[m]/d1[mm]/d2[mm] Tmax [°C] Comment
RESTEC Rtx-CLPesticides 30/0.25/0.25 340 Trace pesticides analysis using ECD
Shimadzu SH-5MS 30/0.25/0.25 330 universal, non-polar (Me/Ph = 95/5 polysiloxane)
Shimadzu 5H-I-1MS 30/0.25/0.25 330 Me 100 polysiloxane
Supelco SLB-5MS 30/0.25/0.25 340 universal, non-polar (Me/Ph = 95/5 polysiloxane)
Supelco SLB-5MS 30/0.32/0.25 340 universal, non-polar (Me/Ph = 95/5 polysiloxane)
Supelco SPB-PUFA 30/0.25/0.20 220 FAMEs, poly(alkylene glycol)
Supleco Equity-1 30/0.32/1.0 325 Me 100 polysiloxane
Agilent HP-88 100/0.25/0.25 250 (88 cyanopropyl)siloxane, FAMEs
Agilent HP-5 30/0.32/0.25 325 universal, non-polar (Me/Ph = 95/5 polysiloxane)
Phenomenex Zebron ZB-1701 30/0.25/0.25 325 (86 cyanopropyl)siloxane
Lion LN-WAX Plus 30/0.25/0.25 250 PEG
l = column length; d1 = outer diameter of the column; d2 = stationary-phase film thickness

 
Key features:

  • Electron impact ionisation (EI) or chemical ionisation (CI) with methane as the reaction gas
  • Direct inlet (DI) capability, maximum desorption temperature approximately 350 °C.
  • Routine scan analyses in the m/z range of approximately 10–1090.
  • Capable of analysing volatile and semi-volatile compounds in solutions of non-polar solvents, as well as performing volatile compound analysis above solid or liquid matrices (heatable up to 200 °C) using the HeadSpace method.
  • NIST spectral library available for common compounds.

Location:
Department of Chemistry, U15/220

Responsible person:
Name:             Robert Vícha
Phone.:           576031103
Mobile:           na
E-mail:             rvicha@utb.cz
Office:             U15/424